Occupational Hazard Study: Measurement of Noise Levels of Dental Equipment used in dental clinic and dental laboratory


Ali H. Otoum BDS, JDB*,Falah H. Hawamdeh BDS, JDB,Ahmed E. Dalalah BDS, JDB***,Amer K. Jodeh BDS, JDB+,Mohammed E. Qatawna BDS, JDB++




ABSTRACT

Objective: To measure noise levels at a dental clinic and a dental lab and compare it with NIOSH recommended exposure level.

 

Methods: Noise level measurements were taken at Prince Rashid Hospital (Royal Medical Services), in both dental clinics and lab. The measurements were done using equipment within these settings with the microphone (amprobe SM-10 sound meter, IEC 651 type 2 ) (Figure 1) located 30cm away from the equipment. Cumulative noise in the clinic and lab was measured by locating the microphone in the centre of the room. All measurements were made through the use of a decibulometer. The average noise level was alculated for each item of equipment and compared with the National Institute for Occupational Safety and Health recommended exposure level (REL) .

 

Results: Values higher than the National Institute for Occupational Safety and Health limit were detected in both the clinics and the lab, with values as high as 98 dB, 93 dB and 85 dB. The highest detected values were for the vacuum, sandblaster, and high-speed handpiece.

 

Conclusion: This study shows that there are high noise levels in both dental clinics and lab. the dental lab has higher noise levels compared with the clinics in terms of the equipment used, as well as cumulative noise level when measured in the center of the lab.

 

Keywords: Noise, Dental equipment, Hearing problems, Dental staff, The Royal Medical Services of Jordan.


RMS April 2021; 28(1): 10.12816/0058881



Introduction 


High noise levels have negative effects on human health, such as increased heart rate, high blood pressure, nervousness and emotional exacerbation (1,2). noise produced in dental clinics results in negative psychological outcomes like anxiety in dental patients and irritation in dentists (2,4). In general, dental clinics use equipment that generate loud noises and sounds. 

Many researchers have studied these levels of noise to understand its effects  on patients and dentists (3–5). The extend of damage caused by equipment noise depends mainly on the intensity and frequency of the sounds and the duration of exposure to the noise (6–8). 

According to the American National Institute for Occupational Safety and Health (NIOSH) and the World Health Organisation, an eight-hour daily occupational exposure to noise is limited to 85 dB(A) as a weighted equivalent sound level (LAeq) for a five-day workweek in any working environment (https://www.cdc.gov/niosh/topics/noise/default.html). NIOSH listed noise-induced hearing loss is the most common occupational injury (10). Many studies have measured noise with the intention of understanding its effects on hearing (6,9,11) or focused on measuring and evaluating the levels of noise as a source of anxiety in children(12). 

The current study focused on the noise in dental clinics and labs. It measured the noise level in dental clinics and at the dental lab of the Royal Medical Service in Jordan (Prince Rashid Hospital).

The aim of this study was to measure noise levels produced by equipment inside dental clinic and dental lab and compare it with the recommended exposure level of NIOSH. The results will be used to understand this problem in this setting, and help with reducing the noise levels if it exceeds the allowed limits. 



METHODS


This study was approved by the Research Ethics Committee of the Royal Medical Services of Jordan. It was conducted in the dental clinics and dental lab of Prince Rashid Hospital, which is part of the Royal Medical Services in Jordan. The researcher used a microphone (amprobe SM-10 sound meter, IEC 651 type 2) (Figure 1) to measure the noise level in decibels (dB), which is the unit of 

Measuring noise in any given setting. The noise of eights items of dental equipment inside the dental clinic and four items of dental equipment in the dental lab was measured, plus the noise of the clinic (cumulative noise level) on a regular working day. (Table I and II)

The microphone used for each measurement was located at a distance of approximately 30 cm away from the equipment, in an attempt to mimic the possible position of the operator (the position of the microphone-holding researcher was maintained for repeated recordings). The measurement was repeated five times for each suggested item of equipment (Table I and II) to calculate maximum noise, minimum noise and average noise produced by each item of equipment in different scenarios. All measurements were recorded with only the required persons inside the clinic or lab, with the doors and windows closed to exclude outside noise and sounds other than the tested equipment.

The measurement also was done inside the clinic by putting the microphone in the centre of the clinic to measure the cumulative noise of the clinic on a regular working day with all daily activities kept as usual; this measurement was repeated five times on different working days to calculate the maximum cumulative noise, minimum cumulative noise and average cumulative noise of the dental clinic. The location of the microphone was the same for all measurements. All measurements were performed with the use of a decibulometer, also known as a sound level meter.



                                                Figure 1 Amprobe SM-10 sound meter, IEC 651 type 2.




RESULTS 


 The measurements of noises in the clinics are shown in Table I. The highest level of noise was recorded for the low speed hand piece during tooth cutting with high volume suction (89.4 dB). The second highest recorded noise level was for the high speed hand piece (89.2 dB).The highest average noise level was also recorded for the high-speed hand piece during tooth cutting with high volume suction (85.5 dB). The second highest average noise level was for the low speed hand piece during tooth cutting with high volume suction (83.8 dB). On the other hand, the lowest recorded nose level was for the micro motor hand piece, and only when it was turned on (62.6 dB).


Noise measurements in the dental laboratory are shown in Table II. The highest recorded level of noise was for the vacuum while it was turned on (100 dB), followed by the sandblaster during sand blasting (97.0 dB). Similarly, the highest average noise was for the vacuum while it was turned on (98.8 dB), followed by the sandblaster (93.5 dB). The highest average recorded noise was 99 dB (Figur 2).



Table I Noise levels of equipment measured in dental clinics

Equipment

Studied process

Min dB level

Max dB level

Average dB

High speed hand piece

turned on only

60.3

66.4

63.35

 

Tooth-cutting without suction

71.6

79.4

75.5

 

Tooth-cutting with low volume suction

75.5

78.8

77.15

 

Tooth-cutting with high volume suction

81.8

89.2

85.5

Low speed handpiece

turned on only

64.6

73.8

69.2

 

Tooth-cutting without suction

69.1

76.0

72.55

 

Tooth-cutting with low volume suction

70.0

73.3

71.65

 

Tooth-cutting with high volume suction

78.2

89.4

83.8

Ultrasonic scaler

 turned on only

58.0

68.0

63

 

Without suction

 

 

 

 

with low volume suction

72.9

75.5

74.2

 

with high volume suction

79.6

83.6

81.6

Micro motor handpiece

turned on only

60.1

65.1

62.6

 

Cutting acrylic

73.3

84.5

78.9

Low volume Suction bump

Running free

69.8

71.7

70.75

 

Touching mucosa

71.4

84.6

78

High volume Suction bump

Running free

71.1

77.1

74.1

 

Touching mucosa

80.5

86.5

83.5

Amalgamator

Turned on

72.9

80.7

76.8

Noise in dental clinic*

 

63.8

77.6

70.7

* From the center of the clinic



Table II Noise levels of equipment measured in dental labs

Equipment

Studied process

Min dB level

Max dB level

Average dB

Stone trimmer

Only turned on

72.1

74.0

73

 

Cutting stone

82.0

86.5

84.2

Sandblaster

Sand blasting

90.0

97.0

93.5

Grinding wheel

Only turned on

68.1

79.1

82.6

 

Cutting acrylic

88.4

80.0

84.2

Vacuum

Turned on

97.5

100.0

98.75

Noise in dental lab*

 

67.4

77.8

72.6





Discussion 


According to NIOSH standards, the average noise level should not exceed 85 dB at any working place for five days working week.(1,13) Constant exposure to noise levels exceeding this limit exposes the workers to an increased risk of hearing loss and tinnitus. This limit is based on the limitations of the human ear and human hearing senses (9). According to the results of this research, the noise averages of dental clinic and lab were high ranging from (62-98 dB). 

Some of these noise levels are higher than NIOSH noise limit (figure 2).

shows a comparison between the recorded average noise levels and the NIOSH limit. The first and second recorded high average noise levels were higher than the NIOSH limit; as the first was 98.8 dB for the vacuum while it was turned on, and the second was 93.5 dB for the sandblaster while it was sand blasting, 

respectively.

It was observed that the highest noise levels were both recorded in the dental laboratory, which had a higher noise levels than the clinics. Both measures were higher than the NIOSH limit, which could indicate a potentially harmful occupational environment for workers inside the lab.

The third highest average measurement was recorded in the clinic, with a value of 85.5 dB. This is also higher than the NIOSH limit, which could lead to negative health consequences for workers at these clinics.

These negative consequences of noise on patient wellbeing are more prevalent in female patients, in the paediatric age group, and in patients living in rural areas (2,4,5,12).

The tools that had the highest noise levels were the vacuum (98.8 dB), sandblaster (93.5 dB), high speed handpiece (85.5 dB), grinding wheel (84.2 dB), low speed handpiece (83.8 dB) and ultrasonic scaler (81.6 dB). Only three of these noise values exceeded the NIOSH limits (over 85 dB). These reported noise levels for dental equipment were similar to levels reported in previous studies (14–17). Qsaibati and Ibrahim reported a noise level for the sandblaster of 96 dB and 92 dB for the handpiece from a 15 cm distance (10), while Al‐Omoush et al. reported in a recently published Jordanian study an amalgamator noise level of 55 dB (3). 

The difference in the recorded amalgamator noise levels between this study and the Al‐Omoush et al. study could be explained by possible differences in equipment type, age and maintenance level. 

For the cumulative noise levels in the clinic and the laboratory recorded by positioning the microphone in the centre of the clinic and the laboratory, measures were 70.7 dB in the clinic and 72.6 dB in the dental lab. The reported clinic noise in this study was slightly higher than those previously reported at the Damascus dental collage and lower than those reported at a Brazilian university dental clinic (8,10). Differences in dental clinics may be due to differences in room size, furniture distribution, equipment, and other factors that could affect noise levels in clinics (8,18). 

In the current study, the reported general noise levels in the dental clinic and lab were under the NIOSH limits; however, these levels could change with the use of equipment (aging and maintenance) and with the number and type of patients in the clinic. It has been reported that paediatric dental clinics have higher noise levels due to children crying and screaming (1,19,20).




Conclusion 


 The noise levels of dental clinics and labs are high, some of them higher than the NIOSH limit of 85 dB which could have harmful consequences on dental care workers in this assessed Jordanian hospital. The dental lab has higher noise levels compared with the clinics in terms of the equipment used, as well as regarding the cumulative noise level when measured in the centre of the lab. 

Further longitudinal research is needed to assess the health effects of these detected high noise levels. It is recommended for dental workers to use hearing protection devices to avoid any possible negative effects of high noise levels, this also valid for dental patients despite short periods of exposure to dental environment.

Raising awareness about hearing precautionary measures, regular maintainers of dental equipment, and regular check-ups at specialised clinics for dental health care providers could be the first step toward a healthy dental occupational environment. 


 


References 



1. Ma KW, Wong HM, Mak CM. Dental Environmental Noise Evaluation and Health Risk Model Construction to Dental Professionals. Int J Environ Res Public Health. 2017 19;14(9). 


2. Yousuf A, Ganta S, Nagaraj A, Pareek S, Atri M, Singh K, et al. Acoustic Noise Levels of Dental Equipments and Its Association with Fear and Annoyance Levels among Patients Attending Different Dental Clinic Setups in Jaipur, India. J Clin Diagn Res. 2014 Apr;8(4):ZC29-34. 


3. Al‐Omoush SA, Abdul‐Baqi KJ, Zuriekat M, Alsoleihat F, Elmanaseer WR, Jamani KD. Assessment of occupational noise‐related hearing impairment among dental health personnel. Jrnl of Occup Health [Internet]. 2020 Jan [cited 2020 Jan 20];62(1). Available from: https://onlinelibrary.wiley.com/doi/abs/10.1002/1348-9585.12093


4. Wannemüller A, Sartory G, Elsesser K, Lohrmann T, Jöhren HP. Modality of fear cues affects acoustic startle potentiation but not heart-rate response in patients with dental phobia. Front Psychol. 2015;6:170. 


5. Ajayi DM, Arigbede AO. Barriers to oral health care utilization in Ibadan, South West Nigeria. Afr Health Sci. 2012 Dec;12(4):507–13. 


6. Alabdulwahhab BM, Alduraiby RI, Ahmed MA, Albatli LI, Alhumain MS, Softah NA, et al. Hearing loss and its association with occupational noise exposure among Saudi dentists: a cross-sectional study. BDJ Open. 2016;2:16006. 


7. Burk A, Neitzel RL. An exploratory study of noise exposures in educational and private dental clinics. J Occup Environ Hyg. 2016 02;13(10):741–9. 


8.  da Cunha KF, dos Santos RB, Klien CA. Assessment of noise intensity in a dental teaching clinic. BDJ Open. 2017 Dec;3(1):17010. 


9. Myers J, John AB, Kimball S, Fruits T. Prevalence of tinnitus and noise-induced hearing loss in dentists. Noise Health. 2016 Dec;18(85):347–54. 


10. Qsaibati ML, Ibrahim O. Noise levels of dental equipment used in dental college of Damascus University. Dent Res J (Isfahan). 2014 Nov;11(6):624–30. 


11.Choi H-J, Hwang T-Y, Jeon M-J. Awareness of occupational hazards and personal protective equipment use among dental hygienists. Yeungnam Univ J Med. 2019;36(1):20–5. 


12.Muppa R, Bhupatiraju P, Duddu M, Penumatsa N, Dandempally A, Panthula P. Comparison of anxiety levels associated with noise in the dental clinic among children of age group 6-15 years. Noise Health. 2013;15(64):190. 


13. Safety NI for O, Health. Criteria for a recommended standard: Occupational noise exposure, revised criteria 1998. US Department of Health and Human Services, Public Health Service, Centers for Disease Control and Prevention. NIOSH Cincinnati, OH; 1998. 


14. Bahannan S, el-Hamid AA, Bahnassy A. Noise level of dental handpieces and laboratory engines. J Prosthet Dent. 1993 Oct;70(4):356–60. 


15. Kadanakuppe S, Bhat PK, Jyothi C, Ramegowda C. Assessment of noise levels of the equipments used in the dental teaching institution, Bangalore. Indian J Dent Res. 2011 Jun;22(3):424–31. 


16. Kulkarni E, Abdallah Y, Hanseman D, Krishnan DG. How Much Noise Is an Oral and Maxillofacial Surgeon Exposed to? J Oral Maxillofac Surg. 2018 Jul;76(7):1400–3. 


17. Sampaio Fernandes JC, Carvalho APO, Gallas M, Vaz P, Matos PA. Noise levels in dental schools. Eur J Dent Educ. 2006 Feb;10(1):32–7. 


18. Ai Z, Mak C, Wong H. Noise level and its influences on dental professionals in a dental hospital in Hong Kong. Building Services Engineering Research and Technology. 2017 Sep;38(5):522–35. 


19. Goswami M, Singh D, Vashist B, Marwaha S. Noise levels and sound pollution associated with various operative procedures and equipments in a pediatric dental environment-A clinical study. J Oral Biol Craniofac Res. 2017 Dec;7(3):182–7. 


20. Jadid K, Klein U, Meinke D. Assessment of noise exposures in a pediatric dentistry residency clinic. Pediatr Dent. 2011 Aug;33(4):343–8. 


21. Voitl P, Sebelefsky C, Mayrhofer C, Woditschka A, Schneeberger V. Noise levels in general pediatric facilities: A health risk for the staff? PLoS ONE. 2019;14(3):e0213722. 


22. Lazar A, Kauer R, Rowe D. Hearing Difficulties Among Experienced Dental Hygienists: A Survey. J Dent Hyg. 2015 Dec;89(6):378–83. 





About
The Journal

The Journal of the Royal Medical Services (JRMS) is an open access journal and it is the official publication for the Royal Medical Services of the Jordanian Armed Forces... Read More

Subscribe to OUR
newsletter

To receive updates on new issues

JRMS Journal

Articles Archive

Archive

Previous Issues

Volume 25
April 2018

Volume 24
December 2017

Volume 24
August 2017

Volume 24
March 2017